Back to top
Send SMS Send Inquiry
+918068095943
Home » Products » New Item » Acetaldehyde, Methanol, & Benzene in CO2 Application

Acetaldehyde, Methanol, & Benzene in CO2 Application

Acetaldehyde, Methanol, & Benzene in CO2 Application
Acetaldehyde, Methanol, & Benzene in CO2 Application
Send Inquiry
Product Code : IMPT
Product Description

Acetaldehyde, Methanol, & Benzene in CO2 Application

The BevAlert Model 8900 provides direct measurement of Acetaldehyde, Methanol, and Benzene (AMB) in Carbon Dioxide.

The instrument is utilized by Specialty Gas Manufacturers and the Food and Beverage Industry to monitor trace impurities in CO2 used in carbonated beverages.

The BevAlert Model 8900 employs a photoionization detector (PID). The AMB in the gas sample are physically separated using proprietary GC columns. A dual-column configuration with timed backflush to vent is used to strip off


Applications:

Measurement of AMB in beverage grade carbon dioxide


Features:

  • Direct measurement of Acetaldehyde, Methanol, and Benzene in Carbon Dioxide
  • Interference free response
  • Automatic baseline adjustment for long-term stability
  • Automatic calibration
  • Analog output ranges are user selectable
  • Concentration and diagnostic alarm relays
  • Graphic display of current or historical concentrations
  • Multipoint sampling options
  • RS-232 and optional LAN

Specifications:

Analysis Time<360 seconds
DetectorPID (High-sensitivity PID Optional)
ColumnCapillary
Oven Temperature100ºC, Nominal
Carrier GasNitrogen, 15 cc/min, Nominal
Lower Detection Limit<50ppb Acetaldehyde, <2ppm Methanol, <2ppb Benzene
Accuracy1% of Full-scale
Precision2% of Measured Value
Span Drift (24HR)<2% of Full-scale
Sample Flow Rate250 – 500 cc/min, typical
OutputAnalog: (5) 0-20ma or 4-20ma loop power supplied, isolated. Selectable for: gas concentration, unintegrated detector signal. Options for up to 16 additional programmable 0-20ma, 4-20ma or voltage outputs: 0-1V, 0-5V, or 0-10V.
Digital: RS-232, optional Local Area Network
Relays(5) User programmable relays for concentration and diagnostic alarms (1A @ 30Vdc). Options for up to 32 additional relays available.
InputsOptional digital input board for 3 contact closure inputs. Supports start analysis, start calibration, and analyze calibration gas sample.